Publication detail

CALIBRATION OF PRECISION ROTARY TABLE USING LASER INTERFEROMETER SYSTEM

ŠINDELÁŘ, M.

Original Title

CALIBRATION OF PRECISION ROTARY TABLE USING LASER INTERFEROMETER SYSTEM

Type

article in a collection out of WoS and Scopus

Language

English

Original Abstract

Calibration of a rotary axis can be useful to determine angular deviations throughout the measuring range. Specified accuracy presented in the specification list represents the worst case, therefore information about magnitudes of deviations of individual measuring points are unknown. The resulting error map of the device under test allows correction of the indicated value. As a result, the measurement accuracy of the device can be increased. Calibration of a precision rotary table was performed using a Keysight 5530 laser interferometer system. The presented paper briefly describes the fundamentals of this portable calibrating system. Also, the results and accuracies are discussed.

Keywords

Calibration;rotary table;laser interferometer

Authors

ŠINDELÁŘ, M.

Released

27. 4. 2021

Location

Brno

ISBN

978-80-214-5942-7

Book

Proceedings I of the 27th Conference STUDENT EEICT 2021

Edition

1st Edition

BibTex

@inproceedings{BUT171610,
  author="Michal {Šindelář}",
  title="CALIBRATION OF PRECISION ROTARY TABLE USING LASER INTERFEROMETER SYSTEM",
  booktitle="Proceedings I of the 27th Conference STUDENT EEICT 2021",
  year="2021",
  series="1st Edition",
  pages="610",
  address="Brno",
  isbn="978-80-214-5942-7"
}