Original Abstract
Calibration of a rotary axis can be useful to determine angular deviations throughout the measuring range. Specified accuracy presented in the specification list represents the worst case, therefore information about magnitudes of deviations of individual measuring points are unknown. The resulting error map of the device under test allows correction of the indicated value. As a result, the measurement accuracy of the device can be increased. Calibration of a precision rotary table was performed using a Keysight 5530 laser interferometer system. The presented paper briefly describes the fundamentals of this portable calibrating system. Also, the results and accuracies are discussed.