Publication detail

DEPENDENCE OF CONTRAST ON PRESSURE USING SEGMENTAL IONIZATION DETECTOR IN ENVIRONMENTAL SEM

JIRÁK, J., ČERNOCH, P., AUTRATA, R.

Original Title

DEPENDENCE OF CONTRAST ON PRESSURE USING SEGMENTAL IONIZATION DETECTOR IN ENVIRONMENTAL SEM

Type

conference paper

Language

English

Original Abstract

Ionization detector utilizes impact ionization in electrostatic field in gaseous environment in the specimen chamber for signal detection. The contrasts between four metal materials using ionization detector with three concentric electrodes in various connection of the electrodes were monitored. As experimentally determined, the grounding of the outer electrode has a positive influence on the image contrast.

Keywords

ionization detector, ESEM

Authors

JIRÁK, J., ČERNOCH, P., AUTRATA, R.

RIV year

2005

Released

24. 2. 2005

Publisher

Česká mikroskopická společnost, Vídeňská 1083, Praha 4

Location

Brno

Pages from

32

Pages to

32

Pages count

1

BibTex

@inproceedings{BUT17218,
  author="Josef {Jirák} and Pavel {Černoch} and Rudolf {Autrata}",
  title="DEPENDENCE OF CONTRAST ON PRESSURE USING SEGMENTAL IONIZATION DETECTOR IN ENVIRONMENTAL SEM",
  booktitle="Mikroskopie 2005",
  year="2005",
  volume="2",
  number="1",
  pages="1",
  publisher="Česká mikroskopická společnost, Vídeňská 1083, Praha 4",
  address="Brno"
}