Přístupnostní navigace
E-application
Search Search Close
Publication detail
PROVAZNÍK, I. NOVÁKOVÁ, M. KOZUMPLÍK, J. VESELÝ, Z. BLAHA, M. CHMELAŘ, M. NOVÁKOVÁ, Z.
Original Title
Acquisition and Pre-Processing Of Action Potentials Optically Recorded From Animal Heart
Type
conference paper
Language
English
Original Abstract
A flexible electro-optical recording system is presented, which allows simultaneous recording of electrograms from orthogonal leads and action potentials from an arbitrary site on heart surface. The recorded action potentials are disturbed with noise from multiple sources. A robust QRS-complex detector with the following modified averaging with filtered residue is proposed. Adding the filtered residue back to the averaged signal helps to keep information on trends in the signal.
Keywords
electro-optical recording system, electrogram, action potential, noise, averaging
Authors
PROVAZNÍK, I.; NOVÁKOVÁ, M.; KOZUMPLÍK, J.; VESELÝ, Z.; BLAHA, M.; CHMELAŘ, M.; NOVÁKOVÁ, Z.
RIV year
2004
Released
1. 6. 2004
Publisher
VUTIUM
Location
Brno
ISBN
80-214-2633-0
Book
Analysis of Biomedical Signals and Images
Edition number
1
Pages from
229
Pages to
231
Pages count
3
BibTex
@inproceedings{BUT11210, author="Valentine {Provazník} and Marie {Nováková} and Jiří {Kozumplík} and Zbyněk {Veselý} and Milan {Blaha} and Milan {Chmelař} and Zuzana {Nováková}", title="Acquisition and Pre-Processing Of Action Potentials Optically Recorded From Animal Heart", booktitle="Analysis of Biomedical Signals and Images", year="2004", number="1", pages="229--231", publisher="VUTIUM", address="Brno", isbn="80-214-2633-0" }