Publication detail

COMPARISON OF THE METHODS USED TO LOW LEVEL DC VOLTAGE MEASUREMENT

OBŠILOVÁ, L.

Original Title

COMPARISON OF THE METHODS USED TO LOW LEVEL DC VOLTAGE MEASUREMENT

Type

article in a collection out of WoS and Scopus

Language

English

Original Abstract

This paper deals with comparison of three methods used to measure low level DC voltage. It is about Reference Step Method, potentiometric method and direct method. Measurement was held in cooperation with the Czech metrology institute. Comparison of the methods consist of a graphic comparison and a key comparison. Results showed good agreement of all methods except for potentiometric method at one range of a device under test.

Keywords

DC voltage; Reference Step Method; potentiometric method; uncertainty

Authors

OBŠILOVÁ, L.

Released

28. 4. 2016

Publisher

Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních

Location

Brno

ISBN

978-80-214-5350-0

Book

Proceedings of the 22nd Conference STUDENT EEICT 2016

Edition

1

Edition number

první

Pages from

406

Pages to

410

Pages count

5

BibTex

@inproceedings{BUT124561,
  author="Lucie {Obšilová}",
  title="COMPARISON OF THE METHODS USED TO LOW LEVEL DC VOLTAGE MEASUREMENT",
  booktitle="Proceedings of the 22nd Conference STUDENT EEICT 2016",
  year="2016",
  series="1",
  number="první",
  pages="406--410",
  publisher="Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních",
  address="Brno",
  isbn="978-80-214-5350-0"
}