Přístupnostní navigace
E-application
Search Search Close
Publication detail
JUREN, T. LÍZAL, F. POHOŘILSKÝ, M.
Original Title
In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates
Type
abstract
Language
English
Original Abstract
Our aim was to compare devices for nasal continuous positive airway pressure (NCPAP) application in neonates by measuring the pressure of the device and flow delivered to the patient. The following NCPAP delivery devices were tested: [Babyflow-Draeger (sizes XS, S, M, L, XL), Vygon single tube 2.5 mm, Vygon double tube 1.5 mm, RAM cannula (for newborns under 1.0 kg), Argyle pharyngeal cannula]. The generated flow was set to 8 liter/min (10 liter/min for L and XL), positive end-expiratory pressure was set to 5 mbar (6 mbar for L and XL).
Keywords
nasal continuous positive airway pressure (NCPAP), neonate, flow
Authors
JUREN, T.; LÍZAL, F.; POHOŘILSKÝ, M.
Released
28. 5. 2017
Location
Westerburg, Germany
ISBN
1934-5798
Periodical
Journal of Neonatal-Perinatal Medicine
Year of study
10
Number
2
State
unknown
Pages from
1
Pages to
Pages count
URL
http://www.jnpm.org/content/abstracts-25th-european-workshop-neonatology-westerburg-castle-germany-may-28th-%E2%80%93-31st-2017
BibTex
@misc{BUT137605, author="Tomáš {Juren} and František {Lízal} and Marek {Pohořilský}", title="In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates", year="2017", journal="Journal of Neonatal-Perinatal Medicine", volume="10", number="2", pages="1--1", address="Westerburg, Germany", doi="10.3233/NPM-990003", issn="1934-5798", url="http://www.jnpm.org/content/abstracts-25th-european-workshop-neonatology-westerburg-castle-germany-may-28th-%E2%80%93-31st-2017", note="abstract" }