Publication detail

Voxel size and calibration for CT measurements with a small field of view

BLAŽEK, P. ŠRÁMEK, J. ZIKMUND, T. KALASOVÁ, D. HORTVÍK, V. KLAPETEK, P. KAISER, J.

Original Title

Voxel size and calibration for CT measurements with a small field of view

Type

journal article - other

Language

English

Original Abstract

Dimensional tomographic measurement is influenced by many factors. To achieve metrological traceability of results, knowledge of influence factors and their effect on measurement is important. This includes knowledge of the true value of voxel size and edge detection. Phantoms used for characterization and correction of error sources are calibrated usually on tactile or optical coordinate measuring machines. The challenge here is the manufacturing and calibration of small phantoms, which can be used in the field of view approximately 1 mm and less, which is the case of tomographic devices aiming at high resolution. In this work, abilities of a nano-coordinate measuring machine (nano-CMM) SIOS NMM-1 (Nanopositioning and Nanomeasuring Machine) to calibration of a phantom for X-ray computed tomography (CT) with the small field of view and high resolution are tested.

Keywords

Radiographic Testing (RT), CT, nano-CMM, SIOS NMM-1, voxel size calibration, edge detection calibration

Authors

BLAŽEK, P.; ŠRÁMEK, J.; ZIKMUND, T.; KALASOVÁ, D.; HORTVÍK, V.; KLAPETEK, P.; KAISER, J.

Released

4. 3. 2019

Publisher

NDT.net

ISBN

1435-4934

Periodical

The e-Journal of Nondestructive Testing

Number

2019

State

Federal Republic of Germany

Pages from

1

Pages to

7

Pages count

7

BibTex

@article{BUT156603,
  author="Pavel {Blažek} and Jan {Šrámek} and Tomáš {Zikmund} and Dominika {Kalasová} and Václav {Hortvík} and Petr {Klapetek} and Jozef {Kaiser}",
  title="Voxel size and calibration for CT measurements with a small field of view",
  journal="The e-Journal of Nondestructive Testing",
  year="2019",
  number="2019",
  pages="1--7",
  issn="1435-4934"
}