Přístupnostní navigace
E-application
Search Search Close
Publication detail
ISLAM, M. KORESHKOVA, A. GUPTA, V. LEWIS, T. MACKA, M. PAULL, B. MAHBUB, P.
Original Title
Fast pulsed amperometric waveform for miniaturised flow-through electrochemical detection: Application in monitoring graphene oxide reduction
Type
journal article in Web of Science
Language
English
Original Abstract
A fast pulsed waveform (2-steps at +400 mV and -400 mV, 10 ms each, 50 Hz cycle repetition) was developed incorporated within a miniaturised EC detector, and used to demonstrate the pulsed amperometric detection (PAD) of hydrazine in a flow-based analytical system. In contrast to standard amperometric detection (AD), a subsequent surface oxide reduction step was applied within the PAD waveform. The PAD applied within a simple flow-injection analysis (FIA) method provided a limit of detection (LOD) just as low as 0.8 nM of hydrazine (RSD 5%, n = 9, linearity r(2) = 0.99 for 1 nM-100 mM), which is one order of magnitude lower than the lowest LOD reported todate. Significantly, the LOD was obtained using 5-20 times lower electrode surface area, flow rate, and sample volume, than in previous methods, with 15 times faster cycle repetition frequency than alternative PAD based studies reported in literature. Practical application of the PAD waveform was demonstrated by monitoring the temporal consumption of reducing agents during graphene oxide reduction experiments, separated using chromatographic methods (hydrazine in IC and ascorbic acid in HPLC).
Keywords
Pulsed amperometric detection and fast waveform; Miniaturised flow-based electrochemical detection; Hydrazine and ascorbic acid; Graphene oxide reduction
Authors
ISLAM, M.; KORESHKOVA, A.; GUPTA, V.; LEWIS, T.; MACKA, M.; PAULL, B.; MAHBUB, P.
Released
20. 12. 2019
ISBN
0013-4686
Periodical
Electrochimica Acta
Year of study
328
Number
1
State
United Kingdom of Great Britain and Northern Ireland
Pages from
Pages to
8
Pages count
URL
https://www.sciencedirect.com/science/article/pii/S0013468619319589
BibTex
@article{BUT160537, author="ISLAM, M. and KORESHKOVA, A. and GUPTA, V. and LEWIS, T. and MACKA, M. and PAULL, B. and MAHBUB, P.", title="Fast pulsed amperometric waveform for miniaturised flow-through electrochemical detection: Application in monitoring graphene oxide reduction", journal="Electrochimica Acta", year="2019", volume="328", number="1", pages="1--8", doi="10.1016/j.electacta.2019.135087", issn="0013-4686", url="https://www.sciencedirect.com/science/article/pii/S0013468619319589" }