Publication detail

Generation and Detection of Hyperpolarized Xenon

RYCHNOVSKÝ, J. BUCHTA, Z.

Original Title

Generation and Detection of Hyperpolarized Xenon

Type

conference paper

Language

English

Original Abstract

We present the experimental results of detection of detection of hyperpolarized Xenon obtained by 4.7 T, 200 MHz magnetic resonance (MR) system. Xenon 129 was hyperpolarized by the spin-exchange with Rubidium pumped by extended cavity laser system based on a tunable high-power laser diode optimized for maximum efficiency of the optical pumping process. The application of the laser system is oriented to employment in an experimental arrangement for production of hyperpolarized gasses (HpG), namely Xenon. It is designed to operate in medical and industrial applications to come.

Keywords

hyperpolarized, 129 Xenon, laser diode system, magnetic resonance, optical pumping

Authors

RYCHNOVSKÝ, J.; BUCHTA, Z.

RIV year

2007

Released

26. 4. 2007

Publisher

VUT v Brně, FEKT a FIT

ISBN

978-80-214-3410-3

Book

Proceedings of the 13th conference Student EEICT 2007

Edition

Volume 4

Edition number

první

Pages from

361

Pages to

365

Pages count

5

BibTex

@inproceedings{BUT22965,
  author="Jan {Rychnovský} and Zdeněk {Buchta}",
  title="Generation and Detection of Hyperpolarized Xenon",
  booktitle="Proceedings of the 13th conference Student EEICT 2007",
  year="2007",
  series="Volume 4",
  number="první",
  pages="5",
  publisher="VUT v Brně, FEKT a FIT",
  isbn="978-80-214-3410-3"
}