New Recording Setup for Ratiometric Recording of Action Potentials by Optical Means

BARDOŇOVÁ, J. PROVAZNÍK, I. NOVÁKOVÁ, M. NOGOVÁ, K. SEKORA, J.

Original Title

New Recording Setup for Ratiometric Recording of Action Potentials by Optical Means

Type

conference paper

Language

English

Original Abstract

Monophasic action potential (MAP) can be recorded from the heart surface by optical method via fluorescence measurement. The motion of isolated heart during experiment caused additional noise in recorded signal. The motion artifact can be eliminated by ratiometric fluorescence emission measurements. This study is based on experiments in which a new ratiometric recording setup is used for recording MAPs from isolated rabbit hearts perfused according to Langendorff. MAPs recorded by new ratiometric recording setup are compared with simple optical MAP measurement without ratiometry.

Keywords

Recording, Ratiometric Recording, Action Potentials

Authors

BARDOŇOVÁ, J.; PROVAZNÍK, I.; NOVÁKOVÁ, M.; NOGOVÁ, K.; SEKORA, J.

RIV year

2008

Released

1. 12. 2008

Publisher

IEEE

Location

Bologna, Italy

ISBN

0-7803-8894-1

Book

Computers in Cardiology

Pages from

1

Pages to

4

Pages count

4

BibTex

@inproceedings{BUT27046,
  author="Jana {Kolářová} and Valentine {Provazník} and Marie {Nováková} and Kateřina {Nogová} and Jiří {Sekora}",
  title="New Recording Setup for Ratiometric Recording of Action Potentials by Optical Means",
  booktitle="Computers in Cardiology",
  year="2008",
  pages="1--4",
  publisher="IEEE",
  address="Bologna, Italy",
  isbn="0-7803-8894-1"
}