Přístupnostní navigace
E-application
Search Search Close
Publication detail
KOUKAL, M. FUIS, V. FLORIAN, Z. JANÍČEK, P.
Original Title
A NUMERICAL STUDY OF EFFECTS OF THE MANUFACTURE PERTURBATIONS TO CONTACTS OF THE TOTAL HIP REPLACEMENT
Type
journal article - other
Language
English
Original Abstract
In the endoprosthesis surgery there is a large percentage of implant defects, which cause the failure of the whole prosthesis. One kind of the total hip replacement functionality loss is acetabular cup pull-off from pelvis bone. This article is aimed at manufacture perturbations analysis as one of the possible reasons to this kind of failure. Dimension and of geometry (roundness) manufacturing perturbations were analyzed. It was found, that these perturbations affect considered values of contact pressure and frictional moment. Contact pressure and frictional moment are quantities affecting replacement success and durability.
Keywords
Total hip replacement, FEM, roundness, contact pressure, frictional moment, biomechanics
Authors
KOUKAL, M.; FUIS, V.; FLORIAN, Z.; JANÍČEK, P.
RIV year
2011
Released
29. 4. 2011
Location
Brno
ISBN
1802-1484
Periodical
Engineering Mechanics
Year of study
18
Number
1
State
Czech Republic
Pages from
33
Pages to
42
Pages count
10
BibTex
@article{BUT76519, author="Milan {Koukal} and Vladimír {Fuis} and Zdeněk {Florian} and Přemysl {Janíček}", title="A NUMERICAL STUDY OF EFFECTS OF THE MANUFACTURE PERTURBATIONS TO CONTACTS OF THE TOTAL HIP REPLACEMENT", journal="Engineering Mechanics", year="2011", volume="18", number="1", pages="33--42", issn="1802-1484" }