Přístupnostní navigace
E-application
Search Search Close
Publication detail
ŠKARVADA, P. TOMÁNEK, P. GRMELA, L. TOFEL, P. ŠEVČÍK, M. MACKŮ, R.
Original Title
Fluctuations of ultrasonic transducer vibration measurement
Type
conference paper
Language
English
Original Abstract
Langevin type transducers are being used by reason of the high transmitting power capability for various purposes. The peak-to-peak value of the transducer mechanical oscillation is in order of tens of the nanometers (for the esonant frequency and two ring transducer). Transducer vibration parameters depend considerably on excitation voltage and frequency. Beyond the resonant frequency some parasitic resonance frequencies are produced. Precise measurement of vibration with nanometer accuracy in the range of the ultrasonic frequencies can be done with using optical interferometry. Results show high distortion of the transducer displacement waveforms and poor amplitude stability.
Keywords
vibrometer; light interference; interferometer; Langevin transducer;
Authors
ŠKARVADA, P.; TOMÁNEK, P.; GRMELA, L.; TOFEL, P.; ŠEVČÍK, M.; MACKŮ, R.
RIV year
2013
Released
28. 6. 2013
Location
Montpellier
ISBN
978-1-4799-0670-3
Book
2013 22nd International Conference on Noise and Fluctuations (ICNF), IEEE Catalog Number CFP1392N-USB
Pages from
Th-P-48-1
Pages to
Th-P-48-4
Pages count
4