Original Abstract
This article describes a designed of a holder for post-mortem XRD measurements of Li-ion, Na-ion and Li-S electrodes, which need to be performed in the inert atmosphere. The first part describes the measurement method and problems arising during the measurements of the samples. Subsequently, the design and construction of a new cell with hermetic insulation and sample displacement error correction possible during the operation of the diffractometer is described. Finally, the inert gas attenuation within the cell is theoretically calculated. The calculation is verified by practical measurements, and alternatives are proposed to improve the signal-to-noise ratio.
Keywords
XRD, ex-situ, post-mortem, sample displacement, attenuation, SNR
Authors
Released
23. 4. 2020
ISBN
978-80-214-5868-0
Book
Proceedings of the 26th Conference STUDENT EEICT 2020 Selected Papers
Edition
1
Edition number
25
Pages from
103
Pages to
107
Pages count
5
BibTex
@inproceedings{BUT166188, author="Ondřej {Klvač}", title="THE DESIGN OF SAMPLE HOLDER FOR XRD MEASUREMENTS IN INERT ATMOSPHERE WITH DISPLACEMENT ERROR CORRECTION", booktitle="Proceedings of the 26th Conference STUDENT EEICT 2020 Selected Papers", year="2020", series="1", number="25", pages="103--107", isbn="978-80-214-5868-0" }