Publication detail

Three-Dimensional Electrogram in Spherical Coordinates: Application to Ischemia Analysis

JANOUŠEK, O. KOLÁŘOVÁ, J. NOVÁKOVÁ, M. PROVAZNÍK, I.

Original Title

Three-Dimensional Electrogram in Spherical Coordinates: Application to Ischemia Analysis

Type

journal article in Web of Science

Language

English

Original Abstract

Three-dimensional electrogram was used for analysis of ischemia manifestation in isolated hearts. Three parameters based on spherical coordination system were used in this study amplitude of electrical heart vector, its azimuth and elevation. The parameters were presented as a trend. This approach reflected ischemic changes in a manner which can be easily observed and evaluated. Ischemia was analysed in seven isolated hearts of New Zealand white rabbits. It was found that (a) ischemia changes heart electric vector, (b) ischemic preconditioning has a protective effect, and (c) both of these findings can be clearly observed by the proposed method.

Keywords

Three dimensional electrogram; Vectorcardiogram; Spatiocardiogram; Vectorcardiography parameters; Myocardial ischemia

Authors

JANOUŠEK, O.; KOLÁŘOVÁ, J.; NOVÁKOVÁ, M.; PROVAZNÍK, I.

RIV year

2010

Released

28. 5. 2010

Publisher

AV ČR

Location

Praha

ISBN

0862-8408

Periodical

Physiological Research

Year of study

59

Number

Suppl 1

State

Czech Republic

Pages from

S51

Pages to

S58

Pages count

8