Ultrasonic transducer peak-to-peak optical measurement

ŠKARVADA, P. TOFEL, P. TOMÁNEK, P.

Original Title

Ultrasonic transducer peak-to-peak optical measurement

Type

journal article - other

Language

English

Original Abstract

Possible optical setups for measurement of the peak-to-peak value of an ultrasonic transducer are described in this work. The Michelson interferometer with the calibrated nanopositioner in reference path and laser Doppler vibrometer were used for the basic measurement of vibration displacement. Langevin type of ultrasonic transducer is used for the purposes of Electro-Ultrasonic Nonlinear Spectroscopy (EUNS). Parameters of produced mechanical vibration have to been well known for EUNS. Moreover, a monitoring of mechanical vibration frequency shift with a mass load and sample-transducer coupling is important for EUNS measurement.

Keywords

Homodyne interferometer, Langevin transducer, light interference, optical vibrometer.