Original Abstract

The paper contains a brief description of the current state in our calibration facility for the acoustic emission sensors. We have implemented the reciprocity method according to the NDIS 2109, ISO/TR 13115, using the step function according to ASTM 1106, ISO 12713 and the longitudinal wave calibration according to NPL. The greatest attention is given to problems relating to the correct implementation of AE sensor calibration methodology. It describes in detail influence of the aperture effect on the frequency response to the Rayleigh wave calibration of a wideband sensor. Briefly described is the basic principle and design of a wideband AE sensor with a conical active element and frequency characteristics of our reference sensors are specified as obtained from all types of absolute calibration for the surface waves and longitudinal waves.

Keywords

acoustic emission, reference sensor, calibration

Authors

BENEŠ, P.; FIALKA, J.

RIV year

2012

Released

12. 9. 2012

Publisher

University of Granada

Location

18071 Granada, Spain

ISBN

978-84-615-9941-7

Book

Proceedings 30th European Conference on Acoustic Emission Testing and 7th International Conference on Acoustic Emission

Edition

1

Edition number

1

Pages from

113-1

Pages to

113-9

Pages count

9