Diffusion MRI: magnetic field inhomogeneities mitigation
MARCOŇ, P. BARTUŠEK, K. GESCHEIDTOVÁ, E. DOKOUPIL, Z.
Original Title
Diffusion MRI: magnetic field inhomogeneities mitigation
Type
journal article - other
Language
English
Original Abstract
The article reports on certain artifacts that emerge during the in vitro diffusion-weighted imaging of physical samples. In this context, the authors analyze the influence of magnetic field inhomogeneity, temperature, or eddy currents and consider procedures to mitigation of the artifacts. A technique reducing the examined spurious effects was designed, experimentally verified, and denominated as the three measurements method. The technique proved to be useful mainly for the evaluation of a DWI image measured with a diffusion gradient in the z axis, where the relative measurement error decreased to 3.38% (during measurement using two images, the relative error was greater than 19%). For small errors within the measurement of diffusion constants of a deionised water sample (<5%) it is necessary to select a b-factor value larger than 200x10^6 s.m-2. Temperature stabilisation with an accuracy better than 0.1 degree of celsius during the entire measuring process is a necessary prerequisite for the measurement of biological or material samples with a relative accuracy lower than 1%.
Keywords
correction, diffusion, inhomogeneity, eddy currents, magnetic resonance
Authors
MARCOŇ, P.; BARTUŠEK, K.; GESCHEIDTOVÁ, E.; DOKOUPIL, Z.
RIV year
2012
Released
17. 9. 2012
Publisher
Versita Open, Versita Ltd. London
Location
London, Great Britain
ISBN
1335-8871
Periodical
Measurement Science Review
Year of study
12
Number
5
State
Slovak Republic
Pages from
205
Pages to
212
Pages count
8
URL
BibTex
@article{BUT93943,
author="Petr {Marcoň} and Karel {Bartušek} and Eva {Gescheidtová} and Zdeněk {Dokoupil}",
title="Diffusion MRI: magnetic field inhomogeneities mitigation",
journal="Measurement Science Review",
year="2012",
volume="12",
number="5",
pages="205--212",
issn="1335-8871",
url="http://www.measurement.sk/2012/Marcon.pdf"
}