Detail publikace

Analysis of the Impact of teh Baffle Placement in the Supersonic Flow in the Differentially Pumped Chamber

BÍLEK, M.

Originální název

Analysis of the Impact of teh Baffle Placement in the Supersonic Flow in the Differentially Pumped Chamber

Typ

článek ve sborníku ve WoS nebo Scopus

Jazyk

angličtina

Originální abstrakt

The thesis follows to the work of G.D. Danilatos occupy Differential Pumping System with fast stopper focusing on dew design for environmental scanning electron microscope. Thesis concern divider located in the chamber differntial pumping and its influence on the nature of the flow. Simulation of individual vairants was evaluated using the program Ansys Fluent, which caluvulates with the method final volumes as opposed to calculations performed by G.D. Danilatos which uses the statistical method Monte Carlo.

Klíčová slova

Supersonic Flow, Ansys Flunet, Differentially Pumped Chamber

Autoři

BÍLEK, M.

Vydáno

27. 8. 2018

Nakladatel

Brno University of Technology

Místo

Brno

ISBN

978-80-214-5651-8

Kniha

Advanced Batteries Accumulators and Fuel Cells – 19th ABAF

Strany od

148

Strany do

152

Strany počet

5

BibTex

@inproceedings{BUT150636,
  author="Michal {Bílek}",
  title="Analysis of the Impact of teh Baffle Placement in the Supersonic Flow in the Differentially Pumped Chamber",
  booktitle="Advanced Batteries Accumulators and Fuel Cells – 19th ABAF",
  year="2018",
  pages="148--152",
  publisher="Brno University of Technology",
  address="Brno",
  isbn="978-80-214-5651-8"
}