Detail publikace

Biomechanical comparison of all-polyethylene total knee replacement and its metal-backed equivalent on periprosthetic tibia using the finite element method

APOSTOLOPOULOS, V. BOHÁČ, P. MARCIÁN, P. NACHTNEBL, L. MAHDAL, M. PAZOUREK, L. TOMÁŠ, T.

Originální název

Biomechanical comparison of all-polyethylene total knee replacement and its metal-backed equivalent on periprosthetic tibia using the finite element method

Typ

článek v časopise ve Web of Science, Jimp

Jazyk

angličtina

Originální abstrakt

BackgroundTotal knee arthroplasty (TKA) with all-polyethylene tibial (APT) components has shown comparable survivorship and clinical outcomes to that with metal-backed tibial (MBT). Although MBT is more frequently implanted, APT equivalents are considered a low-cost variant for elderly patients. A biomechanical analysis was assumed to be suitable to compare the response of the periprosthetic tibia after implantation of TKA NexGen APT and MBT equivalent.MethodsA standardised load model was used representing the highest load achieved during level walking. The geometry and material models were created using computed tomography data. In the analysis, a material model was created that represents a patient with osteopenia.ResultsThe equivalent strain distribution in the models of cancellous bone with an APT component showed values above 1000 mu epsilon in the area below the medial tibial section, with MBT component were primarily localised in the stem tip area. For APT variants, the microstrain values in more than 80% of the volume were in the range from 300 to 1500 mu epsilon, MBT only in less than 64% of the volume.ConclusionThe effect of APT implantation on the periprosthetic tibia was shown as equal or even superior to that of MBT despite maximum strain values occurring in different locations. On the basis of the strain distribution, the state of the bone tissue was analysed to determine whether bone tissue remodelling or remodelling would occur. Following clinical validation, outcomes could eventually modify the implant selection criteria and lead to more frequent implantation of APT components.

Klíčová slova

Total knee arthroplasty; Computational modeling; Finite element method; All-polyethylene tibial component; Metal-backed tibial component; TKR; Knee replacement; FEA

Autoři

APOSTOLOPOULOS, V.; BOHÁČ, P.; MARCIÁN, P.; NACHTNEBL, L.; MAHDAL, M.; PAZOUREK, L.; TOMÁŠ, T.

Vydáno

23. 2. 2024

Nakladatel

BMC

Místo

LONDON

ISSN

1749-799X

Periodikum

Journal of Orthopaedic Surgery and Research

Ročník

19

Číslo

1

Stát

Spojené království Velké Británie a Severního Irska

Strany od

153

Strany do

164

Strany počet

12

URL

Plný text v Digitální knihovně

BibTex

@article{BUT188224,
  author="Vasileios {Apostolopoulos} and Petr {Boháč} and Petr {Marcián} and Luboš {Nachtnebl} and Michal {Mahdal} and Lukáš {Pazourek} and Tomáš {Tomáš}",
  title="Biomechanical comparison of all-polyethylene total knee replacement and its metal-backed equivalent on periprosthetic tibia using the finite element method",
  journal="Journal of Orthopaedic Surgery and Research",
  year="2024",
  volume="19",
  number="1",
  pages="12",
  doi="10.1186/s13018-024-04631-0",
  issn="1749-799X",
  url="https://josr-online.biomedcentral.com/articles/10.1186/s13018-024-04631-0"
}