Detail publikace

Evaluation of EHD model of the inductive flowmeter

FIALA, P. JIRKŮ, T. BĚHUNEK, I.

Originální název

Evaluation of EHD model of the inductive flowmeter

Typ

článek ve sborníku ve WoS nebo Scopus

Jazyk

angličtina

Originální abstrakt

This article deals with physical and chemical processes during measurement with an inductive flowmeter. A theoretical model and an example of numerical solution are presented here. We prepared numerical models based on the combined finite element method (FEM) and the finite volume method (FVM) of one of the three variants and computed the output voltage of the flowmeter electrodes. The model joins the magnetic, the electric and the current fields, the flow field and a chemical nonlinear ion model. The results were obtained by means of the FEM/FVM as a main application in ANSYS software.

Klíčová slova

magnetohydrodynamics, fluid flow measurement, finite element methods, turbulent media, Hall effect

Autoři

FIALA, P.; JIRKŮ, T.; BĚHUNEK, I.

Rok RIV

2007

Vydáno

23. 5. 2007

Nakladatel

Wydrukowano w Zakladzie Graficznym Politechniki Slaskiej w Gliwicach

ISBN

978-83-85940-29-6

Kniha

IC - SPETO 2007

Strany od

63

Strany do

64

Strany počet

2

BibTex

@inproceedings{BUT23688,
  author="Pavel {Fiala} and Tomáš {Jirků} and Ivo {Běhunek}",
  title="Evaluation of EHD model of the inductive flowmeter",
  booktitle="IC - SPETO 2007",
  year="2007",
  pages="63--64",
  publisher="Wydrukowano w Zakladzie Graficznym Politechniki Slaskiej w Gliwicach",