Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
FRK, M.
Originální název
The Accuracy of Dielectric Measurement of Liquid Dielectric Materials
Typ
článek ve sborníku ve WoS nebo Scopus
Jazyk
angličtina
Originální abstrakt
This paper is focused on the accuracy of dielectric measurement of liquid dielectric materials by the dielectric relaxation spectroscopy method in the frequency area. Experimental measurement was realized on the liquid samples of oligobutadiene. The main attention of the experiment is concentrated on the evaluation of the accuracy respectively error and uncertainty of the measurement. Obtained data helps to determine optimum operating conditions of whole measuring system Agilent- precision LCR meter Agilent 4284A and liquid test fixture Agilent 16452A.
Klíčová slova
Accuracy of measurement, Complex permittivity, Liquid dielectric materials, Measuring system Agilent
Autoři
Rok RIV
2007
Vydáno
4. 9. 2007
Nakladatel
Ing. Zdeněk Novotný, CSc.
Místo
Brno
ISBN
978-80-7355-078-3
Kniha
New Trends in Physics - NTF 2007
Číslo edice
1.
Strany od
27
Strany do
30
Strany počet
4
BibTex
@inproceedings{BUT23785, author="Martin {Frk}", title="The Accuracy of Dielectric Measurement of Liquid Dielectric Materials", booktitle="New Trends in Physics - NTF 2007", year="2007", number="1.", pages="27--30", publisher="Ing. Zdeněk Novotný, CSc.", address="Brno", isbn="978-80-7355-078-3" }