Detail publikace
Noise measurement setup for quartz crystal microbalance
SEDLÁK, P. MAJZNER, J. ŠIKULA, J. HÁJEK, K.
Originální název
Noise measurement setup for quartz crystal microbalance
Typ
článek v časopise - ostatní, Jost
Jazyk
angličtina
Originální abstrakt
The paper describes our measurement setup, in which a quartz crystal oscillator with coated active layers and a reference quartz oscillator are driven by two oscillator circuits. Each one regulates a frequency of a crystal at the minimum impedance which corresponds to the series resonance. A data-acquisition card triggers on the rise-edges of the output signal and stores these corresponding times on which the instantaneous frequency is estimated by own-written software.
Klíčová slova
Quartz crystal microbalance, noise measurement setup, fluctuation
Autoři
SEDLÁK, P.; MAJZNER, J.; ŠIKULA, J.; HÁJEK, K.
Rok RIV
2012
Vydáno
12. 4. 2012
Nakladatel
SPOLECNOST PRO RADIOELEKTRONICKE INZENYRSTVI
Místo
Prague
ISSN
1210-2512
Periodikum
Radioengineering
Ročník
21
Číslo
1
Stát
Česká republika
Strany od
207
Strany do
212
Strany počet
6
BibTex
@article{BUT92964,
author="Petr {Sedlák} and Jiří {Majzner} and Josef {Šikula} and Karel {Hájek}",
title="Noise measurement setup for quartz crystal microbalance",
journal="Radioengineering",
year="2012",
volume="21",
number="1",
pages="207--212",
issn="1210-2512"
}