Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
HLAVATÁ, P. MAXA, J. BÍLEK, M. VYROUBAL, P.
Originální název
Impact of the shape of the differentially pumped chamber on critical flow character
Typ
článek ve sborníku mimo WoS a Scopus
Jazyk
angličtina
Originální abstrakt
The article deals with the impact of the shape of the differential pumped chamber of the Environmental electron microscope on resultant value of average Normalized density number on the path of the primary electron beam which has directly proportional influence on the dispersion of this beam and thereby influence on sharpness resulting image. Analysis were made using finite volume method in Ansys Fluent system with the respect to critical flow created behind the aperture which separates differential pumped chamber from specimen chamber.
Klíčová slova
ANSYS Fluent, Electron microscopy, Environmental scanning electron microscope, Dr. Danilatos, Monte Carlo
Autoři
HLAVATÁ, P.; MAXA, J.; BÍLEK, M.; VYROUBAL, P.
Vydáno
11. 9. 2017
Nakladatel
Brno University of Technology
Místo
Brno
ISBN
978-80-214-5529-0
Kniha
Advanced Batteries Accumulators and Fuel Cells - 18th ABAF
Edice
18
Číslo edice
1
Strany od
161
Strany do
163
Strany počet
3
BibTex
@inproceedings{BUT140295, author="Pavla {Šabacká} and Jiří {Maxa} and Michal {Bílek} and Petr {Vyroubal}", title="Impact of the shape of the differentially pumped chamber on critical flow character", booktitle="Advanced Batteries Accumulators and Fuel Cells - 18th ABAF", year="2017", series="18", number="1", pages="161--163", publisher="Brno University of Technology", address="Brno", isbn="978-80-214-5529-0" }