HLAVATÁ, P. MAXA, J. BÍLEK, M. LEPLTOVÁ, K. BAYER, R.
Originální název
Influence of Critical Flow in the Differentially Pumped Chamber AQUASEM
Typ
článek v časopise ve Scopus, Jsc
Jazyk
angličtina
Originální abstrakt
The article deals with the effect of the critical flow, which is produced at pumping of the differential pumping chamber of the Environmental Scanning Electron Microscope (ESEM). The article compares the state of the flow in the area of pumped chamber at pressure ratio on the apertures when the critical flow occurs and its advantages in design of the chambers versus the flow without the clogging the nozzle. Problematics were solved using finite volume method by Ansys Fluent system.
Klíčová slova
ANSYS Fluent, differentially pumped chamber, ESEM, SolidWorks
Autoři
HLAVATÁ, P.; MAXA, J.; BÍLEK, M.; LEPLTOVÁ, K.; BAYER, R.
Vydáno
21. 12. 2017
Nakladatel
University of Defence
Místo
Brno
ISSN
1802-2308
Periodikum
Advances in Military Technology
Číslo
2
Stát
Česká republika
Strany od
301
Strany do
310
Strany počet
10
BibTex
@article{BUT140368, author="Pavla {Šabacká} and Jiří {Maxa} and Michal {Bílek} and Robert {Bayer} and Petr {Vyroubal}", title="Influence of Critical Flow in the Differentially Pumped Chamber AQUASEM", journal="Advances in Military Technology", year="2017", number="2", pages="301--310", doi="10.3849/aimt.01201", issn="1802-2308" }