HRDÝ, R. KYNCLOVÁ, H. KLEPÁČOVÁ, I. BARTOŠÍK, M. NEUŽIL, P.
Originální název
Portable Lock-in Amplifier-Based Electrochemical Method to Measure an Array of 64 Sensors for Point-of-Care Applications
Typ
článek v časopise ve Web of Science, Jimp
Jazyk
angličtina
Originální abstrakt
We present a portable lock-in amplifier-based electrochemical sensing system. The basic unit (cluster) consists of four electrochemical cells (EC), each containing one pseudoreference electrode (PRE) and one working electrode (WE). All four ECs are simultaneously interrogated, each at different frequencies, with square wave pulses superposed on a sawtooth signal for cyclic voltammetry (CV). Lock-in amplification provides independent read-out of four signals, with excellent noise suppression. We expanded a single cluster system into an array of 16 clusters by using electronic switches. The chip with an array of ECs was fabricated using planar technology with a gap between a WE and a PRE of ≈2 μm, which results in partial microelectrode-type behavior. The basic electrode characterization was performed with the model case using a ferricyanide–ferrocyanide redox couple (Fe2+/Fe3+) reaction, performing CV and differential pulse voltammetry (DPV). We then used this system to perform cyclic lock-in voltammetry (CLV) to measure concurrently responses of the four ECs. We repeated this method with all 64 ECs on the chip. The standard deviation of a peak oxidation and reduction current in a single channel consisting of 13 ECs was ≈7.46% and ≈5.6%, respectively. The four-EC configuration in each measured spot allows determination of nonperforming ECs and, thus, to eliminate potential false results. This system is built in a portable palm-size format suitable for point-of-care applications. It can perform either individual or multiple measurements of active compounds, such as biomarkers.
Klíčová slova
CYCLIC VOLTAMMETRY; CANCER DIAGNOSTICS; ELECTRODES; MICROELECTRODE; SYSTEM; CHIP; FABRICATION; CYSTEINE; DESIGN; GOLD
Autoři
HRDÝ, R.; KYNCLOVÁ, H.; KLEPÁČOVÁ, I.; BARTOŠÍK, M.; NEUŽIL, P.
Vydáno
5. 9. 2017
Nakladatel
American Chemical Society, 1155 16TH ST, NW, WASHINGTON, DC 20036 USA
Místo
WASHINGTON
ISSN
0003-2700
Periodikum
ANALYTICAL CHEMISTRY
Ročník
86
Číslo
17
Stát
Spojené státy americké
Strany od
8731
Strany do
8737
Strany počet
7
URL
http://pubs.acs.org/doi/abs/10.1021/acs.analchem.7b00776
BibTex
@article{BUT143063, author="Radim {Hrdý} and Hana {Kynclová} and Ivana {Klepáčová} and Martin {Bartošík} and Pavel {Neužil}", title="Portable Lock-in Amplifier-Based Electrochemical Method to Measure an Array of 64 Sensors for Point-of-Care Applications", journal="ANALYTICAL CHEMISTRY", year="2017", volume="86", number="17", pages="8731--8737", doi="10.1021/acs.analchem.7b00776", issn="0003-2700", url="http://pubs.acs.org/doi/abs/10.1021/acs.analchem.7b00776" }