Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
CHOUDHURY, D. REBENDA, D. SASAKI, S. HEKRLE, P. VRBKA, M. ZOU, M.
Originální název
Enhanced lubricant film formation through micro-dimpled hard-on-hard artificial hip joint: An in-situ observation of dimple shape effects
Typ
článek v časopise ve Web of Science, Jimp
Jazyk
angličtina
Originální abstrakt
This study evaluates the impact of dimple shapes on lubricant film formation in artificial hip joints. Micro-dimples with 20–50 µm lateral size and 1 ± 0.2 µm depths were fabricated on CrCoMo hip joint femoral heads using a picosecond laser. Tribological studies were performed using a pendulum hip joint simulator to apply continuous swing flexion–extension motions. The results revealed a significantly enhanced lubricant film thickness (≥ 500 nm) with micro-dimpled prosthesis heads at equilibrium position after the lubricant film has fully developed. The average lubricant film thickness of dimpled prostheses with square- and triangular-shaped dimple arrays over time is about 3.5 that of the non-dimpled prosthesis (204 nm). Remarkably, the prosthesis with square-shaped dimple arrays showed a very fast lubricant film formation reaching their peak values within 0.5 s of pendulum movement, followed by prosthesis with triangular-shaped dimple arrays with a transition period of 42.4 s. The fully developed lubricant film thicknesses (≥ 700 nm) are significantly higher than the surface roughness (≈ 25 nm) demonstrating a hydrodynamic lubrication. Hardly any scratches appeared on the post-experimental prosthesis with square-shaped dimple array and only a few scratches were found on the post-experimental prosthesis with triangular-shaped dimple arrays. Thus, prostheses with square-shaped dimple arrays could be a potential solution for durable artificial hip joints.
Klíčová slova
Lubricant film formation; Micro-dimple; Artificial hip joints; in-situ; Lubrication
Autoři
CHOUDHURY, D.; REBENDA, D.; SASAKI, S.; HEKRLE, P.; VRBKA, M.; ZOU, M.
Vydáno
1. 5. 2018
Nakladatel
Elsevier B.V.
Místo
Radarweg 29, 1043 NX Amsterdam, The Netherlands
ISSN
1751-6161
Periodikum
Journal of the mechanical behavior of biomedical materials
Ročník
81
Číslo
5
Stát
Nizozemsko
Strany od
120
Strany do
129
Strany počet
10
URL
https://www.sciencedirect.com/science/article/pii/S1751616118301231?via%3Dihub
BibTex
@article{BUT146334, author="CHOUDHURY, D. and REBENDA, D. and SASAKI, S. and HEKRLE, P. and VRBKA, M. and ZOU, M.", title="Enhanced lubricant film formation through micro-dimpled hard-on-hard artificial hip joint: An in-situ observation of dimple shape effects", journal="Journal of the mechanical behavior of biomedical materials", year="2018", volume="81", number="5", pages="120--129", doi="10.1016/j.jmbbm.2018.02.014", issn="1751-6161", url="https://www.sciencedirect.com/science/article/pii/S1751616118301231?via%3Dihub" }