Biomechanical performance of cranial implants with different thicknesses and material properties: A finite element study
Typ
článek v časopise ve Web of Science, Jimp
Jazyk
angličtina
Originální abstrakt
This study investigated the effect of implant thickness and material on deformation and stress distribution within different components of cranial implant assemblies. Using the finite element method, two cranial implants, differing in size and shape, and thicknesses (1, 2, 3 and 4 mm, respectively), were simulated under three loading scenarios. The implant assembly model included the detailed geometries of the mini-plates and micro-screws and was simulated using a sub-modeling approach. Statistical assessments based on the Design of Experiment methodology and on multiple regression analysis revealed that peak stresses in the components are influenced primarily by implant thickness, while the effect of implant material is secondary. On the contrary, the implant deflection is influenced predominantly by implant material followed by implant thickness. The highest values of deformation under a 50N load were observed in the thinnest (1 mm) polymethyl methacrylate implant (Small defect: 0.296 mm; Large defect: 0.390 mm). The thinnest Polymethyl methacrylate and Polyether Ether Ketone implants also generated stresses in the implants that can potentially breach the materials' yield limit. In terms of stress distribution, the change of implant thickness had a more significant impact on the implant performance than the change of Young's modulus of the implant material. The results indicated that the stresses are concentrated in the locations of fixation; therefore, the detailed models of mini-plates and micro-screws implemented in the finite element simulation provided a better insight into the mechanical performance of the implant-skull system.
Klíčová slova
Cranioplasty; Skull implant; 3D printing; Finite element method; Mechanical properties
Autoři
MARCIÁN, P.; NARRA, N.; BORÁK, L.; CHAMRAD, J.; WOLFF, J.
Vydáno
3. 6. 2019
ISSN
0010-4825
Periodikum
COMPUTERS IN BIOLOGY AND MEDICINE
Číslo
109
Stát
Spojené státy americké
Strany od
43
Strany do
52
Strany počet
10
URL
https://www.sciencedirect.com/science/article/pii/S0010482519301234?via%3Dihub
BibTex
@article{BUT156749, author="Petr {Marcián} and Nathaniel {Narra} and Libor {Borák} and Jakub {Chamrad} and Jan {Wolff}", title="Biomechanical performance of cranial implants with different thicknesses and material properties: A finite element study", journal="COMPUTERS IN BIOLOGY AND MEDICINE", year="2019", number="109", pages="43--52", doi="10.1016/j.compbiomed.2019.04.016", issn="0010-4825", url="https://www.sciencedirect.com/science/article/pii/S0010482519301234?via%3Dihub" }