Comparing Ventricular Synchrony in Left Bundle Branch and Left Ventricular Septal Pacing in Pacemaker Patients

HECKMAN, L. LUERMANS, J. ČURILA, K. VAN STIPDONK, A. WESTRA, S. SMÍŠEK, R. PRINZEN, F. VERNOOY, K.

Originální název

Comparing Ventricular Synchrony in Left Bundle Branch and Left Ventricular Septal Pacing in Pacemaker Patients

Typ

článek v časopise ve Web of Science, Jimp

Jazyk

angličtina

Originální abstrakt

Background: Left bundle branch area pacing (LBBAP) has recently been introduced as a novel physiological pacing strategy. Within LBBAP, distinction is made between left bundle branch pacing (LBBP) and left ventricular septal pacing (LVSP, no left bundle capture). Objective: To investigate acute electrophysiological effects of LBBP and LVSP as compared to intrinsic ventricular conduction. Methods: Fifty patients with normal cardiac function and pacemaker indication for bradycardia underwent LBBAP. Electrocardiography (ECG) characteristics were evaluated during pacing at various depths within the septum: starting at the right ventricular (RV) side of the septum: the last position with QS morphology, the first position with r' morphology, LVSP and-in patients where left bundle branch (LBB) capture was achieved-LBBP. From the ECG's QRS duration and QRS morphology in lead V1, the stimulus- left ventricular activation time left ventricular activation time (LVAT) interval were measured. After conversion of the ECG into vectorcardiogram (VCG) (Kors conversion matrix), QRS area and QRS vector in transverse plane (Azimuth) were determined. Results: QRS area significantly decreased from 82 +/- 29 mu Vs during RV septal pacing (RVSP) to 46 +/- 12 mu Vs during LVSP. In the subgroup where LBB capture was achieved (n = 31), QRS area significantly decreased from 46 +/- 17 mu Vs during LVSP to 38 +/- 15 mu Vs during LBBP, while LVAT was not significantly different between LVSP and LBBP. In patients with normal ventricular activation and narrow QRS, QRS area during LBBP was not significantly different from that during intrinsic activation (37 +/- 16 vs. 35 +/- 19 mu Vs, respectively). The Azimuth significantly changed from RVSP (-46 +/- 33 degrees) to LVSP (19 +/- 16 degrees) and LBBP (-22 +/- 14 degrees). The Azimuth during both LVSP and LBBP were not significantly different from normal ventricular activation. QRS area and LVAT correlated moderately (Spearman's R = 0.58). Conclusions: ECG and VCG indices demonstrate that both LVSP and LBBP improve ventricular dyssynchrony considerably as compared to RVSP, to values close to normal ventricular activation. LBBP seems to result in a small, but significant, improvement in ventricular synchrony as compared to LVSP.

Klíčová slova

left bundle branch area pacing; bradycardia pacing; cardiac resynchronization therapy

Autoři

HECKMAN, L.; LUERMANS, J.; ČURILA, K.; VAN STIPDONK, A.; WESTRA, S.; SMÍŠEK, R.; PRINZEN, F.; VERNOOY, K.

Vydáno

17. 3. 2021

Nakladatel

MDPI

Místo

BASEL

ISSN

2077-0383

Periodikum

Journal of Clinical Medicine

Ročník

10

Číslo

4

Stát

Švýcarská konfederace

Strany od

1

Strany do

12

Strany počet

12

URL

Plný text v Digitální knihovně

BibTex

@article{BUT172323,
  author="Luuk {Heckman} and Justin {Luermans} and Karol {Čurila} and Antonius {Van Stipdonk} and Sjoerd {Westra} and Radovan {Smíšek} and Frits {Prinzen} and Kevin {Vernooy}",
  title="Comparing Ventricular Synchrony in Left Bundle Branch and Left Ventricular Septal Pacing in Pacemaker Patients",
  journal="Journal of Clinical Medicine",
  year="2021",
  volume="10",
  number="4",
  pages="1--12",
  doi="10.3390/jcm10040822",
  issn="2077-0383",
  url="https://pubmed.ncbi.nlm.nih.gov/33671420/"
}