Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
BLAŽEK, P. BŘÍNEK, A. VARSLOT, T. ZIKMUND, T. KAISER, J.
Originální název
Effect of detector in-plane rotation misalignment on dimensional measurements using helical CT acquisition
Typ
článek ve sborníku mimo WoS a Scopus
Jazyk
angličtina
Originální abstrakt
One of the critical factors that may significantly contribute to X-ray computed tomography measurement uncertainty is the misalignment of CT system components from ideal geometry assumed by tomographic reconstruction. It is proven to cause image artifacts, inhomogeneous magnification, thus leading to dimensional measurement errors. There is a lack of studies on helical trajectory scanning in metrological applications, focusing on the detector's angular misalignments. We will present results of CT measurements with misaligned detector by in-plane rotation for circular, double-helix, and space-filling trajectory. They prove that such misalignment may cause significant measurements errors, depending on the misalignments magnitude and applied acquisition trajectory.
Klíčová slova
X-ray computed tomography, helical trajectory, space-filling, geometrical misalignments, dimensional measurements
Autoři
BLAŽEK, P.; BŘÍNEK, A.; VARSLOT, T.; ZIKMUND, T.; KAISER, J.
Vydáno
1. 3. 2022
Nakladatel
The e-Journal of Nondestructive Testing
Strany od
1
Strany do
6
Strany počet
URL
http://www.ndt.net/?id=26607
BibTex
@inproceedings{BUT178936, author="Pavel {Blažek} and Adam {Břínek} and Trond {Varslot} and Tomáš {Zikmund} and Jozef {Kaiser}", title="Effect of detector in-plane rotation misalignment on dimensional measurements using helical CT acquisition", year="2022", pages="1--6", publisher="The e-Journal of Nondestructive Testing", url="http://www.ndt.net/?id=26607" }