Originální abstrakt
A method for correction of non-homogenous illumination based on optimization of parameters of B-spline shading model in respect to the Shannons entropy is presented. An analytical gradient term of the entropy criterion has been derived allowing to use the efficient L-BFGS optimization algorithm. The method has been tested on the set of 20 retinal images and has shown ability to substantially eliminate the illumination distortion. The application of this method for reliable image segmentation or registration was tested.
Klíčová slova
illumination correction, optimization, Parzen windowing, retina
Autoři
Rok RIV
2006
Vydáno
26. 6. 2006
Místo
Brno
Strany od
257
Strany do
515
Strany počet
259
BibTex
@inproceedings{BUT18968, author="Libor {Kubečka} and Jiří {Jan} and Radim {Kolář} and Radovan {Jiřík}", title="Retrospective Illumination Correction of Retinal Images", booktitle="Proceedings of 17th International EURASIP Conference BIOSIGNAL 2004", year="2006", volume="17", pages="259", address="Brno" }