3D freehand ultrasound is an imaging technique, which is gradually finding clinical applications. A position sensor is attached to a conventional ultrasound probe, so that B-scans are acquired along with their relative locations. This allows the B-scans to be inserted into a 3D regular voxel array, which can then be visualized using arbitrary-plane slicing, and volume or surface rendering. A key requirement for correct reconstruction is the calibration: determining the position and orientation of the B-scans with respect to the position sensor’s receiver. Following calibration, interpolation in the set of irregularly spaced B-scans is required to reconstruct a regularvoxel array. This text describes a freehand measurement of 2D ultrasonic data, an approach to the calibration problem and several numerical issues concerned with the calibration and reconstruction.
Klíčová slova v angličtině
ultrasonogrpahy, 3D imaging, calibration, positioning
Autoři
Rok RIV
2005
Vydáno
15. 11. 2005
Nakladatel
IFMBE
Místo
Praha
Strany od
222
Strany do
227
Strany počet
6
URL
knihovna UBMI
BibTex
@inproceedings{BUT20438, author="Asterios {Anagnostoudis} and Jiří {Jan}", title="Cross-wire Calibration for Freehand 3D Ultrasonography", booktitle="IFMBE Proceedings (EMBEC)", year="2005", series="IFMBE Proceedings Vol.11, 2005", number="1", pages="6", publisher="IFMBE", address="Praha", url="knihovna UBMI" }