Originální název
Time resolved emission measurement in double-pulse LIBS
Typ
abstrakt
Jazyk
angličtina
Originální abstrakt
For the orthogonal reheating mode, double pulse LIBS experimental setup with ICCD and PMT was designed. The double pulse LIBS plasma is created with a system of two Q switched Nd:YAG lasers. The two lasers (UP-266 MACRO, New Wave and Quantel Brilliant) are operated at the fourth harmonic frequency (266 nm) and at the fundamental wavelength of 1064 nm, respectively. The laser induced plasma radiation is transported by a fiber optic system onto the entrance slit of a monochromator (Jobin Yvon TRIAX 320). ICCD detector (Jobin Yvon Horiba) or a photomultiplier (Hamamatsu) gated by a laboratory built control unit are employed as detectors. For the gate time delay monitoring and time resolved signal recording, a digital storage oscilloscope Tektronix TDS 1012 is utilized.
Klíčová slova
LIBS; time resolved emission measurement
Autoři
Vydáno
7. 7. 2008
Nakladatel
Centre for Geobiology at University of Bergen
Místo
Praha
Strany od
122
Strany do
Strany počet
1
BibTex
@misc{BUT27735, author="Jozef {Kaiser} and Michaela {Vašinová Galiová} and Radomír {Malina}", title="Time resolved emission measurement in double-pulse LIBS", booktitle="9th European Workshop on Laser Ablation in Elemental and Isotopic Analysis", year="2008", pages="122--122", publisher="Centre for Geobiology at University of Bergen", address="Praha", note="abstract" }