Behaviour of the hip joint endoprosthesis ceramic head with manufacturing inaccuracies under ISO 7206-5 loading

FUIS, V., JANÍČEK, P.

Originální název

Behaviour of the hip joint endoprosthesis ceramic head with manufacturing inaccuracies under ISO 7206-5 loading

Typ

článek v časopise - ostatní, Jost

Jazyk

angličtina

Originální abstrakt

The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions. Concerned are shape deflections from the ideal conical surfaces of the stem and the head of the endoprosthesis that were modelled at the macro and micro-level. The problem of stress in the head was solved using the FEM, the Weibull probability model was used for solving the head cohesion failure probability problem. The results of computational modelling are compared with the data obtained by experimental modelling.

Klíčová slova v angličtině

micro-unevenness modelling, stress and failure probability analyses, Weibull weakest link theory

Autoři

FUIS, V., JANÍČEK, P.

Rok RIV

2003

Vydáno

15. 11. 2003

ISSN

1210-2717

Periodikum

Inženýrská mechanika - Engineering Mechanics

Ročník

10

Číslo

5

Stát

Česká republika

Strany od

399

Strany do

411

Strany počet

13

BibTex

@article{BUT41726,
  author="Vladimír {Fuis} and Přemysl {Janíček}",
  title="Behaviour of the hip joint endoprosthesis ceramic head with manufacturing inaccuracies under ISO 7206-5 loading",
  journal="Inženýrská mechanika - Engineering Mechanics",
  year="2003",
  volume="10",
  number="5",
  pages="13",
  issn="1210-2717"
}