Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
ROUBAL, Z. KŘÍŽ, T.
Originální název
Calibration of the apparatus for 3D magnetic measurement for EIT
Typ
článek ve sborníku ve WoS nebo Scopus
Jazyk
angličtina
Originální abstrakt
The paper deals about result measurement for designed apparatus 3D Teslameter magnetic field. This apparatus allows measure direct or alternating magnetic field in environment with strong electromagnetic disturbance. In the case of measure alternating magnetic field is possible measure signal with lower amplitude than electromagnetic disturbance, concrete S/N <0.01. The measurement was performing in workplace for mapping magnetic field. In the designed 3D magnetic field meter, Hall probes are used for the measurement of magnetic flux density; also, the influence of Earths magnetic field is compensated. The measurement result was compared with simulation results in program MATLAB. In this paper are discussing differences between measurement and simulation results and determine source systematic error. The magnetic field map is measured using the calibration 3D meter and is used for the reconstruction of material properties of the measured sample.
Klíčová slova
Hall probe, MIT,EIT
Autoři
ROUBAL, Z.; KŘÍŽ, T.
Rok RIV
2012
Vydáno
18. 4. 2012
Nakladatel
The Electromagnetic Academy
Místo
Cambridge
ISBN
978-1-934142-20-2
Kniha
Proceedings of PIERS 2012 in Kuala Lumpur
Strany od
385
Strany do
389
Strany počet
5
URL
http://piers.org/piersproceedings/piers2012KualalumpurProc.php?searchname=Mikulka
BibTex
@inproceedings{BUT92019, author="Zdeněk {Roubal} and Tomáš {Kříž}", title="Calibration of the apparatus for 3D magnetic measurement for EIT", booktitle="Proceedings of PIERS 2012 in Kuala Lumpur", year="2012", pages="385--389", publisher="The Electromagnetic Academy", address="Cambridge", isbn="978-1-934142-20-2", url="http://piers.org/piersproceedings/piers2012KualalumpurProc.php?searchname=Mikulka" }