WANDROL, P.
Original Title
Úhlové rozložení zpětně odražených elektronů v ESEM.
English Title
Angular distribution of backscattered electrons signal in Environmental Scanning Electron Microscopy
Type
conference paper
Language
Czech
Original Abstract
Cílem práce bylo ověřit vliv prostorového úhlu sběru zpětně odražených elektronů na velikost detekovaného signálu v ESEM.
English abstract
The aim of this work was to verify an influence of the collection solid angle on a magnitude of detected signal. The presented results of measurement was gained by an observation of the specimen of carbon covered with a thin layer of gold. The primary beam accelerating voltage was 20 kV. The collection solid angle was changed by application of aluminium masks with aperture diameters 2, 3, 4, 8, 16 mm mounted on scintillator. As a working environment in the specimen chamber were used air and saturated water vapour.
Keywords
prostorový úhel. zpětně odražené elektrony, ESEM
Key words in English
solid angle, backscattered electrons, ESEM
Authors
RIV year
2004
Released
23. 10. 2002
Publisher
Západočeská univerzita v Plzni
Location
Plzeň
ISBN
80-7082-904-4
Book
Elektrotechnika a informatika 2002
Edition number
1
Pages from
325
Pages to
328
Pages count
4
BibTex
@inproceedings{BUT13049, author="Petr {Wandrol}", title="Úhlové rozložení zpětně odražených elektronů v ESEM.", booktitle="Elektrotechnika a informatika 2002", year="2002", number="1", pages="4", publisher="Západočeská univerzita v Plzni", address="Plzeň", isbn="80-7082-904-4" }