The scintillator-photomultiplier detection system installed under the pole piece represents the most efficient BSE detector in SEM. However, the initial energy of the BSEs of 0.5-3 keV in the LV SEM is the energy, on which the light yield of scintillators decreases. Therefore, it is necessary to accelerate the BSEs energy by electrostatic field before they reach the scintillator. By this electrostatic field both BSEs and SEs are accelerated to the scintillator and the image is a mixture of material and topographical contrasts. It is necessary to separate SEs by an energy filter or an immersion magnetic lens for the true material contrast observation.