MRAVCOVÁ, L. FUČÍK, J.
Original Title
CHEMICAL EVALUATION OF MEDICAL DEVICES (DENTAL COMPOSITES) PART 1
Type
report
Language
English
Original Abstract
Chemical characterization of extracts from two dental composite materials (FD4X Fibrafill Dentin and ever X Posterior) was carried out. Materials were extracted into three extraction liquids (water, buffer – PBS, 100% ethanol) at two different temperatures (37 °C and 50 °C). The LC-MS analysis proved that dental material from ADM, a.s. is equally biocompatible to the reference material from GC EUROPE N.V. in extractability of substances and therefore their biocompatibility should be similar as well.
Keywords
dental composites, monomers, substances of photoinitiator system,extraction, diffusion, liquid chromatrography, mass spectrometry