Publication detail

Primary Calibration of Acoustic Emission Sensors

KEPRT, J.

Original Title

Primary Calibration of Acoustic Emission Sensors

Type

conference paper

Language

English

Original Abstract

The paper reviews the background, the methodology and the standardization of the primary calibration of acoustic emission sensors. Two basic calibration methods, which were practically realized on our test stand, are closely described and explained. It is the method of reciprocity calibration according to [NDIS 2109, 1991] and the step function calibration according to [ASTM 1106, 1986]. As a transfer medium for wave propagation a large steel block was used. The comparison of the results of the both method is presented.

Keywords

acoustic emission, sensor calibration, ASTM 1106, NDIS 2109

Authors

KEPRT, J.

RIV year

2007

Released

24. 5. 2007

Publisher

TU Košice, BERG Faculty

Location

Košice

ISBN

978-80-8073-805-1

Book

Proceedings of 8th international Carpathian Control Conference

Edition number

1

Pages from

287

Pages to

290

Pages count

4