Přístupnostní navigace
E-application
Search Search Close
Publication detail
Musilová Věra, Dupák Jan, Hanzelka Pavel, Králík Tomáš, Urban Pavel
Original Title
Economical helium bath cryopump: design and testing
Type
journal article - other
Language
English
Original Abstract
A small UHV helium bath cryopump with low cryoliquid consumption has been designed, manufactured and tested. The cryopump is designed to keep UHV in electron optical devices without generation of disturbing electromagnetic fields or vibrations. The outer volume of its shell is 15 l, the filling volumes of liquid helium (LHe) and liquid nitrogen (LN2) are 3.0 l and 3.4 l, respectively. Operating times between cryoliquid fillings are 35 days for LHe and 6 days for LN2. A pump speed of 25 l/s was measured for the He gas. A lowest pressure of 1.5,10,7 Pa was achieved in a test chamber and the pumping tests indicated that an even lower pressure would be achieved in a well-degassed chamber. Good agreement between the calculated and measured values both of the pump speed and the cryoliquid evaporation has been found. The system of chevron-type baffles was designed and checked by Monte Carlo simulations of molecular flow and radiative heat transfer. The pump prototype was successfully used to evacuate an operating cryostat of a NMR spectrometer.
Key words in English
Cryopump; Molecular flow; Radiative heat transfer; Cryogenic optimisation; Monte Carlo particle tracing
Authors
Released
3. 5. 2004
ISBN
0042-207X
Periodical
Vacuum
Year of study
2004
Number
74
State
United Kingdom of Great Britain and Northern Ireland
Pages from
77
Pages to
83
Pages count
7
URL
http://www.sciencedirect.com/science?_ob=ArticleURL&_udi=B6TW4-4BHV85H-8&_user=2654390&_handle=V-WA-A-W-AZ-MsSAYWW-UUW-U-AAWCUCYBCW-AAWWZBEACW-WCUBAEAWY-AZ-U&_fmt=full&_coverDate=05%2F03%2F2004&_rdoc=8&_orig=browse&_srch=%23toc%235552%232004%23999259998%23496357!&_cdi=5552&view=c&_acct=C000058444&_version=1&_urlVersion=0&_userid=2654390&md5=8c43f958caf972caf17a35aceb2e3423
BibTex
@article{BUT42434, author="Pavel {Urban}", title="Economical helium bath cryopump: design and testing", journal="Vacuum", year="2004", volume="2004", number="74", pages="7", issn="0042-207X", url="http://www.sciencedirect.com/science?_ob=ArticleURL&_udi=B6TW4-4BHV85H-8&_user=2654390&_handle=V-WA-A-W-AZ-MsSAYWW-UUW-U-AAWCUCYBCW-AAWWZBEACW-WCUBAEAWY-AZ-U&_fmt=full&_coverDate=05%2F03%2F2004&_rdoc=8&_orig=browse&_srch=%23toc%235552%232004%23999259998%23496357!&_cdi=5552&view=c&_acct=C000058444&_version=1&_urlVersion=0&_userid=2654390&md5=8c43f958caf972caf17a35aceb2e3423" }