Originální abstrakt
This paper compares a linear interpolation, an interval and a sine interpolation method for measuring mechanical shock duration. All methods are analysed to find differences and limitations. Then methods are tested on simulated and real shock shapes to confirm analysis results. We discov- ered that the sine interpolation method is very robust to noise, however is sensitive to shock shape changes. The linear interpolation method accuracy is influenced by noise, however this method works also on high noise level (SNR = 0 dB). The interval method provides consistent results up to some noise level (SNR > 10 dB), where it stops working.
Klíčová slova
mechanical shock, half-sine shock, duration measurement, linear interpolation method, interval method, sine interpolation method
Autoři
Vydáno
27. 4. 2017
Nakladatel
Vysoké učení technické v Brně
Místo
Brno
ISBN
978-80-214-5496-5
Kniha
Proceedings of the 23nd Conference STUDENT EEICT 2017
Edice
1
Číslo edice
Strany od
436
Strany do
440
Strany počet
5
BibTex
@inproceedings{BUT135147, author="Jan {Kunz}", title="COMPARISON OF METHODS FOR MEASURING SHOCK DURATION", booktitle="Proceedings of the 23nd Conference STUDENT EEICT 2017", year="2017", series="1", number="1", pages="436--440", publisher="Vysoké učení technické v Brně", address="Brno", isbn="978-80-214-5496-5" }