KOUKAL, M.
Originální název
ANALYSIS OF THE MANUFACTURE PERTURBATIONS OF THE TOTAL HIP REPLACEMENT
Typ
článek ve sborníku ve WoS nebo Scopus
Jazyk
angličtina
Originální abstrakt
One type of total hip replacement function loss is acetabular cup loosening from the pelvic bone. This article examines manufacture deviations as one of the possible reasons for this kind of failure. Both dimensions and geometry manufacturing deviations of ceramic head and polyethylene cup were analyzed. We find that deviations in the variables analysed here affect considered values of contact pressure and frictional moment. Furthermore, contact pressure and frictional moment are quantities affecting replacement success and durability.
Klíčová slova
Total hip replacement, FEM, roundness, contact pressure, frictional moment
Autoři
Vydáno
15. 5. 2017
Nakladatel
VUT v Brně
Místo
Brno
ISBN
978-80-214-5497-2
Kniha
Engineering Mechanics 2017
Číslo edice
ročník 23
ISSN
1805-8248
Periodikum
Engineering Mechanics ....
Ročník
23
Stát
Česká republika
Strany od
506
Strany do
509
Strany počet
4
BibTex
@inproceedings{BUT156140, author="Milan {Koukal}", title="ANALYSIS OF THE MANUFACTURE PERTURBATIONS OF THE TOTAL HIP REPLACEMENT", booktitle="Engineering Mechanics 2017", year="2017", journal="Engineering Mechanics ....", volume="23", number="ročník 23", pages="506--509", publisher="VUT v Brně", address="Brno", isbn="978-80-214-5497-2", issn="1805-8248" }