Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
HARABIŠ, V. KOLÁŘ, R. MÉZL, M. JIŘÍK, R.
Originální název
Comparison and evaluation of indicator dilution models for bolus of ultrasound contrast agents
Typ
článek v časopise ve Web of Science, Jimp
Jazyk
angličtina
Originální abstrakt
Dynamic contrast-enhanced ultrasound (DCE-US) imaging is a promising diagnostic method, which enables the evaluation of tissue perfusion via different parameters. The mean transit time and time-to-peak parameters are the main time parameters and their values depend on the model used for the approximation of the noisy perfusion curves. In this paper, we described a new comparison of different perfusion models using a tissue mimicking phantom. The following models were compared: log-normal, lagged, Erlang, Gamma and the local density random walk model. We discovered that the mean-square error is not the best criterion for model evaluation. More important is the comparison between the estimated time perfusion parameters and the physical parameters of the developed tissue mimicking phantom. Based on the statistical analysis, we can suggest that for the DCE-US perfusion analysis more models should be used, excluding the log-normal model, which gives the highest error of mean transit time value.
Klíčová slova
Ultrasonography, Contrast agent, Perfusion analysis, Perfusion models
Autoři
HARABIŠ, V.; KOLÁŘ, R.; MÉZL, M.; JIŘÍK, R.
Rok RIV
2013
Vydáno
26. 1. 2013
ISSN
0967-3334
Periodikum
PHYSIOLOGICAL MEASUREMENT
Ročník
34
Číslo
2
Stát
Spojené království Velké Británie a Severního Irska
Strany od
151
Strany do
162
Strany počet
12
BibTex
@article{BUT97515, author="Vratislav {Harabiš} and Radim {Kolář} and Martin {Mézl} and Radovan {Jiřík}", title="Comparison and evaluation of indicator dilution models for bolus of ultrasound contrast agents", journal="PHYSIOLOGICAL MEASUREMENT", year="2013", volume="34", number="2", pages="151--162", issn="0967-3334" }