Přístupnostní navigace
E-přihláška
Vyhledávání Vyhledat Zavřít
Detail publikace
BLAŽEK, P. BŘÍNEK, A. VARSLOT, T. ZIKMUND, T. KAISER, J.
Originální název
Effect of detector in-plane rotation misalignment on dimensional measurements using helical CT acquisition
Typ
prezentace, poster
Jazyk
angličtina
Originální abstrakt
Misalignments of X-ray computed tomography system's components are one of the most critical factors causing dimensional measurement uncertainty. In this work we analyzed what dimensional errors are induced by the detector in-plane rotation. The effects were also evaluated and compared between helical and circular trajectory. We demonstrated how can these errors be eliminated using a self-calibration algorithm, which uses only projections data itself without additional reference scanning or precise hardware alignment.
Klíčová slova
X-ray computed tomography, helical trajectory, space-filling, geometrical misalignments, dimensional measurements
Autoři
BLAŽEK, P.; BŘÍNEK, A.; VARSLOT, T.; ZIKMUND, T.; KAISER, J.
Vydáno
17. 5. 2021
Místo
London, United Kingdom
BibTex
@misc{BUT175355, author="Pavel {Blažek} and Adam {Břínek} and Trond {Varslot} and Tomáš {Zikmund} and Jozef {Kaiser}", title="Effect of detector in-plane rotation misalignment on dimensional measurements using helical CT acquisition", year="2021", address="London, United Kingdom", note="presentation, poster" }